CliniCool© Ultra Series 12 Cu. Ft. NSF Certified Pharmacy/Vaccine Refrigerator | Solid Door

Item#: LHU-12-SD-PHNSF

The CliniCool Ultra series NSF certified solid door medical refrigerators are designed specifically for the storage of a variety of vaccines, pharmaceuticals and other temperature sensitive materials often used within the clinical and medical industries. These cutting-edge NSF certified pharmacy refrigerators are certified in accordance with the NSF/ANSI 456 Standard for Vaccine Storage. With this certification, units protect pharmaceuticals at optimal temperatures, preventing waste and allowing for peak delivery. These medical grade refrigerators provide superior sample security with audible & visual high/ low temperature alarms and remote alarm contacts for complete temperature control and awareness. The LHU-12-SD-PHNSF solid door vaccine refrigerators utilize microprocessor controllers and feature temperature alarms, remote alarm contacts, and probe access ports with included probes. Units run on natural, hydrocarbon refrigerant for environmental health and energy efficiency.

 

Includes: Temperature Monitor Device - Complies with The Current CDC Guidelines, 3 Years Certification Of Calibration, “Buffered” Probe In The Product Simulated Solution, Min/Max Memory, °F/°C Switchable, Field Installable, And Visual & Audible Temperature Alarms

What does NSF certified vaccine storage mean? Click here to learn more

$5,736.00 $4,416.72

Request a Quote
  • Temperature setpoint range: +1°C to +10°C (Controller settings must remain unaltered to ensure thermal performance compliant with NSF/ANSI 456 Standard for Vaccine Storage requirements)
  • Operational environment: Indoor use only, +18°C to +26°C (+65°F to +78°F), <70% RH
  • Storage capacity: 12 cu. ft. gross volume
  • Exterior Dimensions (WxDxH): 25" x 29" x 65 3/4"
  • Interior Dimensions (WxDxH): 20 3/8" x 19 1/4" x 36 1/8"
  • Door: One swing solid door, self-closing, right hinged, non-reversible, magnetic sealed gasket, keyed lock
  • Shelves: Four shelves (three adjustable/one fixed) with guard rail on back
  • Mounting: 3 1/2" Swivel Casters (two locking)
  • Interior lighting: Shielded, switched LED lighting, full coverage, balanced spectrum
  • Airflow management: Forced Air technology, patent pending
  • External probe access: Rear wall port (3/4") dia.
  • Insulation: Cabinet is foamed-in-place with EPA compliant high density urethane foam
  • Exterior materials: White powder coated steel
  • Access control: Pyxis®, Omnicell® and AcuDose RX® compatible
  • General warranty: One (1) years parts and labor warranty, excluding display probe calibration
  • Compressor warranty: Five (5) years compressor warranty
  • Product Weight: 231 lbs.
  • Shipping Weight: 256 lbs.
  • Rated Amperage: 3 Amps
  • Power Plug/Power Cord: NEMA 5-15 plug, 8 to 10 ft typical, conforms to UL471 requirements, Vaccine storage power cord warning label
  • Facility Electrical Requirement: 110-120V AC: 15 A (minimum)
  • Agency Listing and Certification: Certified with the temperature performance requirements as defined in the NSF/ANSI 456 Standard for Vaccine Storage for all testing scenarios. UL, C-UL, ETL, C-ETL listed and certified to UL471 standard, hydrocarbon refrigerant safety.
  • Included Accessories:
    • Pharmacy refrigerator/freezer toolkit and temperature logs
    • Temperature Monitor Device - Complies with The Current CDC Guidelines, 3 Years Certification Of Calibration, “Buffered” Probe In The Product Simulated Solution, Min/Max Memory, °F/°C Switchable, Field Installable, And Visual & Audible Temperature Alarms

 

Performance
  • Uniformity¹ (Cabinet air): +/- 0.7°C
  • Stability² (Cabinet air): +/- 1.3°C
  • Maximum temperature variation (Cabinet air): +/- 1.1°C
  • Temperature rise after an after 8 sec door openings: Temperature did not exceed 5.4°C at any probe for all required NSF/ANSI 456 testing protocols
  • Recovery after 3 min door opening: All probes recover to under 8°C within 2.5 min.
  • Energy consumption: 1.32 KWh/day
  • Average heat rejection: 2.21 KWh/day (315 BTU/h)
  • Noise pressure level (dBA): 49 or less installed
  • Pull down time to nominal operating temp: 30 min

Note: The CDC requires a data logger or temperature monitoring device to consistently record live temperature readings from the inside of a refrigerator or freezer when storing all types of vaccines. Constant monitoring of vaccines at recommended temperatures is essential for maintaining the efficacy of a vaccine. Failing to do so can lead to wasted vaccines or ineffective vaccines being administered to patients. Please follow the CDC requirements for Vaccine Storage & the COVID-19 Vaccination Program found here.

Controller, Configuration, Alarms and Monitoring
  • Controller technology: Parametric, microprocessor, LED display with 0.1°C resolution
  • Display Technology: NSF/ANSI 456 Standard for Vaccine Storage compliant digital temperature display and alarm module with battery back-up, F/C switchable.
  • Temperature setpoint range: +1°C to +10°C (Controller settings must remain unaltered to ensure thermal performance compliant with NSF/ANSI 456 Standard for Vaccine Storage requirements)
  • Display probe: Calibrated, stainless steel
  • External alarm connection: 
    • State switching remote alarm contacts.
    • Visual and audible indicators.
  • Alarms: High / Low temperature, compliant with alarm requirements defined in the NSF/ANSI 456 Standard for Vaccine Storage
  • Simulator ballast: Glass bead thermal media
Refrigeration System
  • Compressor: Hermetic, high performance
  • Refrigerant: EPA SNAP compliant, R290, propane
  • Condenser: Fin and tube design, high efficiency fan
  • Evaporator: Fin and tube design, high efficiency fan
  • Defrost: Cycle optimized, zero energy

LHU-12-SD-PHNSF Datasheet

Download

LabRepCo CliniCool NSF Certified Refrigerator & Freezer Owners Manual