The CliniCool© Silver Series pharmacy refrigerator and freezer combination units are designed with a variable speed compressor to deliver unparalleled temperature control. Safeguard your samples with superior temperature uniformity and recovery, even after frequent door openings. Monitor your medical equipment though intuitive digital displays and protective alarms. Champion sustainability with natural hydrocarbon refrigerants—which not only contribute to energy savings, but also foster a greener footprint. Elevate your medical cooling experience with an integrated solution that brings together innovation, performance, and precision.

$9,226.00 $7,104.02

Request a Quote
  • Cu. Ft.: 16 (13 Ref./3 Freezer)
  • Door(s): Double Swing Solid Right Hinged Doors
  • Temperature Range: Refrigerator: +1°C to +10°C; Freezer:-15°C to -25°C
  • Exterior Dimensions: 31"W x 35 1/2"D x 76"H; Door swing: 32 1/2"; Total open Depth: 62 1/4"
  • Interior Dimensions (Freezer): 23 3/4"W x 13 1/4"D
  • Interior Dimensions (Refrigerator): 26"W x 19 3/4"D
  • Audible and visual high and low temperature alarms, refrigerator and freezer, 36 hour battery back-up on alarms, high/low temperature & loss of power, remote alarm contacts, door alarm
  • Casters: n/a
  • Shelves: 3 adjustable, vented stainless steel shelves refrigerator / 1 fixed shelf freezer
  • Interior lighting: 12V LED light bar (Refrigerator)
  • Forced draft air circulation (Refrigerator)
  • Probe access port (3/4") dia., 2 ea. (1 for Refrigerator/1 for Freezer)
  • Cabinet is foamed-in-place with EPA compliant high density urethane foam
  • Exterior: Bioscience Grey powder coated canopy, white textured chamber exterior
  • Leveling Legs
  • Keyed Door Lock
  • Electrical Requirements:  110-120V AC: 15 A (minimum); 3 Amps
  • NEMA 5-15 Plug/ Vaccine storage power cord warning label
  • ETL, C-ETL listed and certified to UL 471 Standard for Commercial Refrigerators and Freezers and CSA C22.2#120 Refrigeration Equipment
  • Product Weight: 262 lbs.
  • Shipping Weight: 291 lbs.
  • Warranty: Two (2) years parts and labor warranty; Seven (7) years compressor warranty

Performance:

  • Refrigerator Uniformity¹ (Cabinet air): +/-0.22°C
  • Refrigerator Stability² (Cabinet air): ±0.48°C
  • Refrigerator Maximum temperature variation (Cabinet air): ±0.71°C
  • Freezer Uniformity¹ (Cabinet air): ±0.69°C
  • Freezer Stability² (Cabinet air): ±1.14°C
  • Freezer Maximum temperature variation (Cabinet air): ±1.95°C
  • Energy Consumption (KWh/day): 3.73
  • Average Heat Rejection (BTU/hr): 5.25
  • Average Heat Rejection (BTU/hr): 746
  • Noise Pressure Level (dBA): 48 or less installed
  • Pull down time to nominal operating temp (Frg/Frz): 40 min / 165 min
  • Variable speed (VSC), Refrigerator. Rated speed range: 1300-4000 rpm
  • Refrigerant: R600a
  • Condenser: Anti-fouling tube and grid design
  • Evaporator: Fin and tube design, high efficiency fan
  • Auto defrost freezer / cycle defrost refrigerator
  • Individual XRi77CH microprocessor temperature controllers for both the refrigerator and freezer with digital temperature display, °C/°F switchable
  • Bi-Colored LED display and alarm module with .1°C resolution, touchpad controls, resettable 1-point min/max history
  • Adjustable operating temperature range: Refrigerator: +1°C to +10°C; Freezer:-35°C to -15°C
  • Digital Communication: RS-485 (MODBUS)
  • Chart Recorder: N/A
  • Battery back-up: 12V high-capacity battery, controller, all alarms active, temperature monitoring DAQ and event logging active on battery backup
  • State switching remote alarm contacts
  • Audible and visual high and low temperature alarms, refrigerator and freezer, 36 hour battery back-up on alarms, high/low temperature & loss of power, remote alarm contacts, door alarm
  • 2 probes for each chamber; 1 in air and 1 in sample bottle

**Notes on Performance Data: Performance data acquired at 22°C ambient, 5°C/-33°C nominal set points in an empty cabinet with shelves using air probes, during stabilized steady state operation and a DAQ sampling rate of one measurement every 10 seconds

1 - Uniformity is defined as the maximum variance in temperature across all probes at any point in time over the testing period
2 - Stability is defined as the maximum variance in temperature experienced by any single probe over the testing period
3 - Data per Energy Star test results or equivalent testing and calculation. Heat rejection based on daily averages, not continuous operation. Performance exceeds Energy Star requirements.
4- Charts are representative of the product and actual performance may vary slightly

LHP-16-RFCAHC-PH Datasheet

Download

Request a Proposal