Battery test chambers from BINDER are ideal for aging, performance, and stress testing in quality assurance and battery production. To ensure maximum safety, all the chambers in the LIT MK series are fitted with detection and fire suppression equipment as standard. TÜV SÜD AG has recently confirmed that the LIT MK 240 and LIT MK 720 battery test chambers are suitable for EUCAR LEVEL 6.

Benefits:

  • Safety thanks to extensive protective measures
  • Homogeneous climate conditions thanks to APT.line™ preheating chamber
  • Extensive programming and data collection
  • Excellent overview of the test object thanks to large, heated viewing window
Request a Quote
Supplier:

Performance Data Temperature

  • Temperature range: -40-100°C
  • Temperature uniformity dependent on set value [± K]: 0.1-0.5
  • Temperature fluctuation dependent on set value [± K]: 0.1-0.5
  • Average heating-up rate according to IEC 60068-3-5: 5.0 K/min
  • Cooling down time from 110°C to -40°C: 160 min
  • Average cooling down time according to IEC 60068-3-5: 3.5 K/min

Electrical data

  • Rated Voltage [V]: 400
  • Power frequency [Hz]: 50
  • Nominal power [kW]: 5.6
  • Unit fuse [A]: 16A
  • Phase (Nominal voltage): 3~

Measures

  • Interior volume [cu.ft.] 8.05/228 L
  • Net weight of the unit (empty) [lbs]: 826.73
  • Permitted load [lbs]: 154.32
  • Load per rack [lbs]: 66.14
  • Wall clearance back [in]: 11.81
  • Wall clearance sidewise [in]: 7.87

Internal Dimensions

  • Width [in]: 28.94
  • Height [in]: 27.56
  • Depth [in]: 17.44
  • Viewing window width [in]: 20
  • Viewing window height [in]: 11.81
  • Unit Doors: 1

Housing dimensions (not including fittings & connections)

  • Width net [in]: 0.04
  • Height net [in]: 0.04
  • Depth net [in]: 36.42

Environment-specific data

  • Sound-pressure level: 62dB(A)
  • Energy consumption at 20°C: 1,300 Wh/h

Fixtures

  • Number of shelves (std./max.): 1/16
  • Temperature range: -40 °C to +110 °C
  • Class 2 independent temperature safety device when temperature is set to +120 °C
  • Reversible stainless steel pressure relief flap installed in the middle on top of the unit
  • Enhanced door-locking mechanism with reinforced brackets
  • Monitoring of CO, H2, O2, and temperature in testing area atmosphere
  • Inertization equipment
  • CO2 fire suppression
  • APT.line™ preheating chamber technology
  • Programmable condensation protection for samples
  • Heated viewing window with LED interior lighting
  • BINDER Multi Management Software APT-COM™ basic edition
  • Intuitive touchscreen controller with time-segment and real-time programming
  • Internal data logger, measured values can be read out in open format via USB
  • Access port with silicone plug, 50 mm, left
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • Computer interface: Ethernet
  • 1 stainless steel rack, can be freely positioned
  • CFC-free refrigerant R-452A
  • Cooling with compressor cooling unit
  • TÜV SÜD confirms that the LIT MK 240 and LIT MK 720 chambers are suitable for EUCAR level 6