800-521-0754

Request Information

CLOSE

Accuris AutoMATE™ 96 Channel Pipetting Station, 115V

Item#: MP9604
  • Fast and accurate 96-channel pipetting
  • Accepts 96/384 well microplates, PCR plates and reservoirs
  • Touch-screen tablet for programming & operation
  • 4-position motorized platform
  • Exchangeable pipette heads (1-20µL, 5-200µL, 100-1000µL)

*Head not included

$15,800.00 $15,010.00

Request a Quote
  • Display: 8-inch color touch screen tablet (included)
  • Aspirate/Dispense volume: 1-1000µl (depending on pipetting
  • Platform capacity: 4 (tip racks/plates/reservoirs)
  • Channels: 96 individual pistons
  • Pipetting technology: Air Displacement
  • Compatible plate formats: 96/384 well microplates, shallow/deep well plates, PCR plates
  • Communication Ports: USB Ports for importing/exporting
  • Voltage Input: Universal 120 to 240VAC, 50-60Hz
  • Power: 100W
  • Weight: 32.6 kg / 71.8 lbs
  • Dimensions: 69.0 x 35.0 x 67.3 cm / 27.2 x 13.8 x 26.5 in.

 

Pipetting Head Specifications
  • Accuracy:
    • MP9600-20PH (1-20µL)
      • ≤12 % at 1μL
      • ≤5 % at 2 μL
      • ≤1 % at 10μL
      • ≤1 % at 20μL
    • MP9600-200PH (5-200µL)
      • ≤5 % at 5μL
      • ≤2 % at 20μL
      • ≤1 % at 100μL
      • ≤1 % at 200μL
    • MP9600-1000PH (100-1000µL)
      • ≤2.5 % at 100μL
      • ≤0.4 % at 500μL
      • ≤1 % at 1000μL
  • Precision:
    • MP9600-20PH (1-20µL)
      • ≤120% at 1μL
      • ≤5 % at 2 μL
      • ≤1 % at 10μL
      • ≤0.8 % at 20μL
    • MP9600-200PH (5-200µL)
      • ≤3.5 % at 5μL
      • ≤1.5 % at 20μL
      • ≤0.8 % at 100μL
      • ≤0.4 % at 200μL
    • MP9600-1000PH (100-1000µL)
      • ≤1.25 % at 100μL
      • ≤0.4 % at 500μL
      • ≤0.4 % at 1000μL
Pipetting Modes:

Standard Pipetting: Aspirating and dispensing reagents by hand is labor-intensive and error prone. With the AutoMATE 96, reagents can be aspirated and dispensed automatically, ensuring accurate and precise transfer to and from microplates or reservoirs.

Multi-Dispensing (Repeat Pipetting): Continually filling microplates is a time-consuming process that can often result in hand fatigue. With the Multi-Dispensing function, repeatedly filling microplates can be automated and accomplished quickly without introducing hand strain or fatigue.

Secondary-Dispensing (Plate Filling): Many protocols and applications call for filling a plate with multiple samples or reagents. With the AutoMATE 96, multiple reagents can be aspirated and dispensed into the same microplate. For improved results, each dispense can be programmed to be include a mixing step.

Serial Dilutions: Performing serial dilutions via manual methods is labor-intensive and error prone. Errors in this process will amplify subsequent dilutions as the process continues. With the AutoMATE 96, serial dilutions can be performed automatically and error free.

Mixing:  Mixing reagents by hand is strenuous and can be inconsistent. With the AutoMATE 96, mixing can be performed automatically with adjustable speeds, ensuring consistency.

Reverse Pipetting: When the accuracy of pipetting volume is critical, the AutoMATE 96 can be set to reverse pipetting mode. This technique greatly improves accuracy &precision with small volumes of biological fluids.

 

Software and Programming

Two operating modes are available and can be selected per the user’s preference.

Manual Mode: Perform operations quickly & efficiently simply by selecting each function individually.

Manual functions:
• Load tips
• Eject tips
• Aspirate reagents
• Dispense reagents
• Mix reagents
• Blow out reagents

Automated Program Mode: Create a protocol made up of multiple steps in succession. This interface allows for complex protocols to be run from tip-loading through completion with the touch of a button.

Programmable steps:
• Load/Eject Tips
• Aspirate/Dispense reagents
• Perform serial dilutions
• Rinse tips
• Repeat a cycle
• Pause after a certain step

Broad Labware Compatibility: All brands of microplates and reservoirs meeting SLAS/ANSI formats are compatible for use with the AutoMATE 96. Included plate adapters allow 384 well microplates and PCR plates to be used.

Fully Automated Tip Loading and Ejection: Motorized tip loading and ejection provides handsfree, consistent tip loading throughout all 96 wells.

Accuris Microplate Instruments Brochure

Download

Accuris AutoMATE 96 Microplate Pipetting Station